Fuente de Poder 4 canales 384 W Rohde & Schwarz HMP4040

Fuente de Poder 4 canales 384 W Rohde & Schwarz HMP4040

El producto no esta disponible en nuestro inventario, consulte con ejecutivo el tiempo de disponibilidad antes de comprar
Vendedor
Rohde & Schwarz
Precio habitual
$ 122,247.83
Precio de oferta
$ 90,553.95

Características y nombre Ficha Técnica
Fuente de Poder 4 canales 384 W Rohde & Schwarz HMP4040
Outputs All channel outputs are galvanically isolated and not connected to ground.
Number of output channels 4
Total output power max. 384 W
Maximum output power per channel 160 W
Output voltage per channel 0 V to 32 V
Maximum output current per channel 10 A
Maximum voltage in serial operation 128 V
Maximum current in parallel operation 40 A
Voltage ripple 3 Hz to 100 kHz < 250 μV (RMS), typ. <150 μV (RMS)
Voltage ripple 3 Hz to 20 MHz typ. < 1.5 mV (RMS)
Current ripple < 1 mA (RMS)
Load regulation load change from 10 % to 90 %  
Voltage ±(% of output + offset) < 0.01 % + 2 mV
Current ±(% of output + offset) < 0.01 % + 250 μA
Load recovery time to within ± 10 mV of the set nominal voltage < 1 ms
Line regulation ±10 % change in mains voltage  
Voltage ±(% of output + offset) < 0.01 % + 2 mV
Current ±(% of output + offset) < 0.01 % + 250 μA
Programming resolution
Voltage 1 mV
Current < 1 A: 0.2 mA; ≥ 1 A: 1 mA
Programming accuracy
Voltage ±(% of output + offset) < 0.05 % + 5 mV
Current ±(% of output + offset) < 0.1 % + 5 mA
Output measurements
Measurement functions voltage, current
Readback resolution
Voltage 1 mV
Current < 1 A: 0.2 mA; ≥ 1 A: 1 mA
Readback accuracy
Voltage ±(% of output + offset) < 0.05 % + 5 mV
Current ±(% of output + offset) < 0.1 % + 2 mA
Temperature coefficient (per °C) 5 °C to 20 °C and 30 °C to 40 °C  
Voltage ±(% of output + offset) 0.01 % + 2 mV
Current ±(% of output + offset) 0.02 % + 3 mA
Sense function yes, for each channel
Maximum sense compensation 1 V
Ratings
Maximum voltage to earth 150 V (peak)
Maximum counter-voltage voltage with same polarity connected to the outputs 33 V
Maximum reverse voltage voltage with opposite polarity connected to the outputs 0.4 V
Maximum reverse current for 5 minutes max. 5 A
Remote control mode
Command processing time < 50 ms
Protection functions
Overvoltage protection configurable for each channel
Overcurrent protection (electronic fuse) configurable for each channel
Response time (Iload > Iresponse × 2) < 10 ms
Fuse linking (FuseLink function) yes
Fuse delay time configurable for each channel 0 ms to 250 ms (10 ms increments)
Overtemperature protection yes
Special functions
Arbitrary function (EasyArb)
Parameter voltage, current, time
Maximum number of points 128
Dwell time 10 ms to 60 s
Repetition continuous or burst mode
1 to 255 repetitions
Data memory non-volatile memory for three arbitrary functions and ten instrument settings
Display and interfaces
Display 240 pixel × 128 pixel LCD
Front panel connections channel outputs 4 mm safety sockets
Rear panel connections   connector block with 4 lines per channel
Remote control interfaces standard dual interface LAN/USB
optional dual interface USB/RS-232 (R&SHO720)
optional IEEE488 (GPIB) (R&SHO740)
General data
Environmental conditions
Temperature operating temperature range +5 °C to +40 °C
storage temperature range -20 °C to +70 °C
Humidity noncondensing 5 % to 80 %
Power rating
Mains nominal voltage 115 V/230 V (±10 %); CAT II
Mains frequency 50 Hz to 60 Hz
Maximum power consumption 600 W
Mains fuses (115 V power source) 2 × T10H/250 V
Mains fuses (230 V power source) 2 × T5H/250 V
Product conformity
Electromagnetic compatibility EU:
in line with EU EMC Directive 2014/30/EU
in line with EU EMC Directive 2014/53/EU





Korea
applied harmonized standards:
❙  EN 61326-1: 2013
❙  EN 61326-2-1: 2013
❙  EN 55011: 2016 (Class A)
❙  IEC 61000-3-2 2014 (Class B)
❙  IEC 61000-3-3: 2013
❙  EN 50581:2012 (RoHS)
❙  EN 300 328 V2.1.1
❙  KN 61000-4-11: 2008
Electrical safety EU:
in line with Low Voltage Directive 2006/95/EC
applied harmonized standards:
EN 61010-1:2010
USA, Canada CSA C22.2 No. 61010-1:2012
Mechanical resistance
Vibration sinusoidal 5 Hz to 55 Hz, 0.3 mm amplitude const.,
55 Hz to 155 Hz, 0.5 g const.,
in line with EN 60068-2-6: 2008
  random 8 Hz to 500 Hz, 1.2 g (RMS), in all 3 axes,
in line with EN 60068-2-64: 2008
Mechanical data
Dimensions (W × H × D) 285 mm × 136 mm × 405 mm (11.22 in × 5.35 in × 15.94 in)
Weight 12.8 kg (28.2 lb)
Rack installation R&S HZP91 option